The maintenance center of Xi'an Antai Test Equipment Co., Ltd. has various types of instrument maintenance business. The repairable brands include: Tektronics, keysight, fluke, Keithley, Rhode and Schwartz, lecory, Anritsu, IFI / Ar / Bonn power amplifier maintenance, Stanford / Newton instrument maintenance, and other brands of instrument or spectrometer maintenance.
Recently, Agitek Service received a Yokogawa WT2030 digital power meter. The customer reported that the instrument failed to connect to a large current and burned. Please listen to the editor of the power meter repair center to explain the case of Yokogawa's digital power meter repairing WT2030 that cannot be turned on. The engineer detects the fault: the instrument has obvious burn marks, the instrument cannot be turned on after power-on, and the power module is burnt out after the test.
Recently, the customer sent an E4438C signal source, and the fault is manifested as self-checking error 411,413; below 4G, the error is unlevel, the output signal is about 10dB lower, and the highest can only go up to -10dB; the signal above 4G is 2~3dB higher. After testing, the output board of the instrument was damaged, and the composite board was damaged, resulting in an error and abnormal output amplitude. Next, the engineer will carry out maintenance processing, replace the damaged components of the output board, replace the damaged components of the composite board, and adjust and test the whole machine. The instrument is repaired, the self-test is ok, and the test is normal. After quality inspection and verification of performance indicators, it will be released from the warehouse.
Feilier infrared thermal imager, T650sc / T630sc series infrared thermal imager is economical, can provide thermal image and visible light image, the resolution can be accurate to point temperature, and the temperature measurement accuracy is very reliable. The built-in digital camera, voice annotation, laser target locator, GPS and other functions are favored by technical experts, engineers and researchers. The rotatable infrared assembly is flexible, allowing you to conduct experiments quickly and comfortably. Repair case of abnormal focus of FILR T630 thermal imager! 1. Maintenance case of FILR T630 thermal imager focusing abnormally 1. Model: T630
Recently, Agitek Service received an electronic load from the customer. After receiving it, the engineer tested the instrument and found that the electronic load could not be loaded. The following will share the maintenance process with you. Model: CH9710C Fault: customer reported for repair (current does not display, may be open circuit). Detection: It can not be uploaded but can be turned on, and the power supply part is excluded. Check out the power section next...
According to the editor of Agitek RF signal source maintenance network, the self-test error of the Keysight signal generator is a common fault in the signal generator. No, Agitek repair center recently received a Keysight E8257D signal generator, and the customer described the failure as a self-check error. The following is a sharing of the maintenance process of this instrument by Liu Gong from the Agitek RF Source Maintenance Center. The engineer of the Agitek RF source repair center inspected the instrument according to the fault described by the customer. After the inspection, it was found that the instrument sampling board and the frequency divider board were damaged. Keysight E8257D signal generator self-checking and error reporting maintenance case will be repaired by engineers of Agitek RF source maintenance center
Agitek Service various types of instrument maintenance services, including oscilloscope maintenance, instrument maintenance, spectrum analyzer maintenance, power supply maintenance, network analyzer maintenance, signal source maintenance, calibration source maintenance, thermal imager maintenance, etc. Sales of testing equipment, video testing equipment, environmental testing equipment, teaching testing equipment, and electronic components; integration of measurement and control systems and monitoring systems; sales and development of computer software, hardware and network equipment; maintenance of instruments (except measuring instruments).
Recently, I received a Keithley 2510 source meter sent by a customer in Shenzhen for repair. The customer described that the source meter could not read the temperature of the load after it was connected to the load. Let's take a look at the causes of damage to this instrument and the repair process.
Recently, I received a Keithley 2510 source meter sent by a customer in Guangzhou for repair. When the customer described the test temperature of the source meter, the error was large. Let's take a look at the causes of damage to this instrument and the repair process.
Recently, a Rohde & Schwarz R&S ZNB20 network analyzer sent by a customer in Shenzhen was repaired. The customer described that the baseline of the instrument was uneven. Let's take a look at the causes of damage to this instrument and the repair process.
Recently, a Keysight DSA-X93304Q oscilloscope sent by a customer in Guangdong was repaired. The customer described that the instrument would not be turned on. Let's take a look at the causes of damage to this instrument and the repair process.
Some time ago, a Keysight DSO9404A oscilloscope was repaired because it failed to turn on. Recently, another one was connected. The customer feedback is that the touch screen cannot be used normally. Let's follow the engineer of the Agitek oscilloscope repair team to understand the specific repair process of this instrument: failure phenomenon : The touch screen cannot be used normally after the instrument is turned on...
This week, a customer sent a Keysight MSO9254A oscilloscope for repair. The customer reported that the oscilloscope can be turned on normally, but it will automatically power off and shut down after a period of use. The following Agitek oscilloscope repair center shares the simple repair process of this instrument: Inspection process: start the inspection after receiving the instrument, first verify the customer's report...
Recently, a Keysight E3646A power supply sent by a customer in Xi'an was repaired. The customer described that the instrument could not output. Let's take a look at the causes of damage to this instrument and the repair process. Trouble phenomenon The instrument cannot output. Detection process After power-on, it can be turned on, but it is found that the voltage cannot be adjusted. By using...
Recently, a Keysight N5182B signal generator sent by a customer in Jiangsu was repaired. The customer described that the instrument would not be turned on. The following N5182B repair center editor will take you to see the cause of damage to this instrument and the repair process.
Recently, an Agilent/Keysight N5183A signal generator sent by a customer in Shenzhen was repaired. The customer described that the instrument suddenly short-circuited and tripped during standby when it was turned on. Later, it was found that the tail had a burnt smell, and now the instrument cannot be turned on. Let's take a look at the causes of damage to this instrument and the repair process. Troubleshooter...
Recently, Agitek Instrument Repair Center has repaired an N9030A spectrum analyzer. Let’s take a look at the simple repair process of this instrument: 1. Fault phenomenon: the attenuation power of 10db and above is not allowed; ...
小黄片软件下载測試維修中心分享是德N5182A信號發生器不同故障維修,一起來看看維修過程吧。N5182A維修小黄片软件下载維修在對是德各型號信號源維修過程中,總結了安捷倫信號源常見的故障問題可供大家分享: 是德信號源常見的故障問題:不開機、開機無顯示、反複重起或死機、開機報錯、自檢報錯、按鍵不靈或失靈、GPIB通訊不良、進不去係統、係統中病毒,花屏、屏幕拖屏、頻率偏差、輸出功率低、無輸出、頻率不準等、計量指標不合格/不通過、輸出端口損壞等等一係列問題。 為避免信號源發生故障,請按照儀器使用注意事項使用信號源、定期對信號源進行保養。 本文由西安小黄片软件下载信號發生器維修
Recently, an Agilent/Keysight N5181A signal generator sent by a customer in Nanjing was repaired. The customer described that the instrument would not be turned on. N5181A maintenance Let's take a look at the cause of damage to this instrument and the maintenance process.
Infrared thermal imager is a high-performance detection technology, but it is inevitable that some faults will occur in the process of use, which will affect the detection results. Ti25 maintenance we must maintain a high degree of attention, usually pay attention to use and maintenance, try to avoid such a situation, so as not to adversely affect our work. Infrared thermal imaging...
Recently, Agitek repaired a Keysight network analyzer E8363C from a university in Beijing. Before E8363C repair, it repaired the problem of loss of lock, no boot problem, and irregular restart of the same type of instrument. E8363C repaired this time. The repaired fault is that the screen has a phase lock error after booting. The following Agitek network maintenance center shares the instrument...
When the Keysight E4445A spectrum analyzer is being repaired, if measuring a high-power signal greater than 30dBm, it is best to add an attenuator for testing, so as not to burn out the spectrum analyzer due to excessive power; 2. The left side of the spectrum analyzer is the display screen, and the right side is the operation button. On the bottom left is the switch. The operation on the right is divided into 5 parts: FUNCION, MARKE, SYSTEM, CONTROL, DATAENTEY. When a key is selected, the corresponding menu option will appear on the right side of the display screen, and the corresponding operation can be selected by pressing the key next to it. 3. When the Keysight E4445A spectrum analyzer is repaired, the input port of the spectrum analyzer has a maximum allowable safe power input, so the maximum input signal cannot exceed the maximum input level allowed by the spectrum analyzer.
Agitek Service Spectrum Analyzer Repair Center, what should I do if the spectrum analyzer is broken? Is the maintenance process and technology of the spectrum analyzer difficult? N9010A spectrum analyzer repair case Yesterday, Agitek Service spectrum analyzer repair center received a N9010A spectrum analyzer sent by the customer. According to the customer's introduction, the fault was that the N9010A could not receive signals. The customer guessed that the incoming DC signal might have burned. Spectrum analyzer N9010A cannot receive signal maintenance case
I believe that engineers are not unfamiliar with network analyzers. Have you encountered network analyzer failures in daily use? Agitekservice will share with you the common failures during network analyzer maintenance. 1. Power failure When the spectrum analyzer is repaired, the power failure is usually manifested as: the power indicator on the front panel does not light up, the fan does not rotate, the instrument cannot be started normally, or the power supply voltage is unstable, causing the spectrum analyzer to keep resetting, and the analog voltage is low. The signal power is low, etc. 2. Display failure Display failure is usually manifested as: no display (black screen), garbled characters, etc.
Xi 'an Agitek Test Equipment Maintenance Co., LTD., founded in 2008, is engaged in test and measurement instrument maintenance and technical services, with foreign high-end instrument chip level maintenance ability, has more than 70 professional maintenance equipment, and professional automatic calibration software. Support Tektronix, KEYSIGHT, Agilent, FLUKE, KEITHLEY, R&S and Anrtisu and other dozens of foreign brands test instrument maintenance. Agitek instrument maintenance center in line with the concept of "honesty and trustworthiness, service first", to serve the majority of electronic test instrument users in China and the Asia-pacific region. The accumulative service customer more than ten thousand, repairs each kind of instrument ten thousand.